The operation of these instruments is based on measuring light absorbance changes resulting from arterial blood flow pulsations. These oximeters include red and infrared light sources typically LEDs , photodetectors, and probes that transmit light through a pulsating arterial bed, such as the fingertip, earlobe, or toe. A co-oximeter, on the other hand, is invasive and uses a blood sample to measure a different value: SaO2, which represents fractional oxygen saturation.
This is the ratio of oxygenated hemoglobin to total hemoglobin. The SaO2 measurement takes into account dysfunctional hemoglobin hemoglobin that cannot carry oxygen, also called dyshemoglobin , which the pulse oximeter measurement does not. Co-oximeters provide A RMS accuracy is a statistical calculation of the difference between device measurements and reference measurements.
Masimo devices with SpCO monitoring are not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; they are intended to be used in conjunction with additional methods of assessing clinical signs and symptoms. Caution: Federal USA law restricts this device to sale by or on the order of a physician. See instructions for use for full prescribing information, including indications, contraindications, warnings, and precautions.
Submit a request to receive more information and a Masimo representative will contact you shortly. Carbon Monoxide Carbon Monoxide. Pulse CO-oximetry is a continuous and noninvasive method of measuring the levels of various blood constituents, including carbon monoxide SpCO.
Hence oxygen saturation is not the sole indicator of oxygen transport. For many purposes, s O 2 either measured by pulse oximeter or calculated by a blood gas analyzer is sufficient to make clinical decisions. When applied appropriately, pulse oximetry can result in advantages such as continuous monitoring, reduced cost and reduced blood loss important when caring for neonates.
However, when poisoning by carbon monoxide or other substances that can affect hemoglobin is suspected, F O 2 Hb, as measured with a bench hemoximeter, is required. Clinical guidelines for the use of pulse oximeters and hemoximeters are available from the American Association for Respiratory Care [6,7].
Related recommendations have been published by the National Committee for Laboratory Standards [8,9]. May contain information that is not supported by performance and intended use claims of Radiometer's products. See also Legal info. Printed from acutecaretesting. June To co-ox or not to co-ox. History The study of blood oxygen saturation has its roots in early hot-air and hydrogen balloon flights made in France during the s [1].
How to measure oxygen saturation There are two basic ways to measure hemoglobin oxygen saturation in blood: 1 gasometrically and 2 spectrophotometrically. Total hemoglobin, in principle, includes all types of hemoglobin: Hemoglobin HbA - normal adult hemoglobin is a complex protein containing iron and capable of transporting oxygen in the blood.
Oxyhemoglobin O 2 Hb - oxygenated hemoglobin, containing four molecules of oxygen per hemoglobin molecule. Carboxyhemoglobin COHb - hemoglobin bound to carbon monoxide, a bond about times stronger than the oxygen-hemoglobin affinity; prevents normal transfer of oxygen and carbon dioxide in the blood. Methemoglobin MetHb - hemoglobin molecule whose iron is in the oxidized, ferric state; useless for respiration; found in the blood after poisoning with acetanilide, potassium chlorate and other substances.
Sulfhemoglobin - hemoglobin in combination with sulphur. The very rare and non-oxygen-carrying sulfhemoglobin is not included in the reported ctHb. Fetal hemoglobin HbF - the major type of hemoglobin in the developing fetus. The oxygen dissociation curve for fetal hemoglobin is shifted to the left compared to adult hemoglobin.
Reference ranges c tHb a reference range adult : Male: 8. It is calculated as follows: The systematic symbol for arterial blood is F O 2 Hb a. Clinical application Oxygen content is a key indicator of oxygen transport within the body. Oxygen transport, defined as the amount of oxygen being transported per liter arterial blood, depends primarily on: The total content of oxygen in the arterial blood, c tO 2 - the key parameter for evaluation of oxygen transport The concentration of hemoglobin in the blood ctHb The concentration of dyshemoglobins c COHb and c MetHb The arterial oxygen tension p O 2 The arterial oxygen saturation s O 2 , which again is determined by p O 2 and p 50 Hence oxygen saturation is not the sole indicator of oxygen transport.
Conclusion For many purposes, s O 2 either measured by pulse oximeter or calculated by a blood gas analyzer is sufficient to make clinical decisions. Astrup P, Severinghaus JW. The history of blood gases, acids and bases. Copenhagen: Munksgaard, The determination of gases in blood and other solutions by vacuum extraction and manometric measurement. J Biol Chem ; Tobin MJ. Principles and practice of intensive care monitoring.
New York: McGraw-Hill, The effect of measured versus calculated hemoglobin oxygen saturation, carboxyhemoglobin and methemoglobin on the pulmonary shunt calculation. Respir Care ; Variation in the hemoglobin-oxygen dissociation curve in arterial blood samples. American Association for Respiratory Care.
Clinical practice guideline: In vitro pH and blood gas analysis and hemoximetry. Clinical practice guideline: Pulse oximetry. Reference and selected procedures for the quantitative determination of hemoglobin in blood.